| E0 Calibration Range | 
1 – 5 * kg  | 
| E1 Calibration Range | 
1 – 5 kg  | 
| E2 Calibration Range | 
1 – 5 kg  | 
| F1 Calibration Range | 
1 – 5 kg  | 
| F2 Calibration Range | 
1 – 5 kg  | 
| Maximum capacity [Max] | 
5.05 kg  | 
| Readability [d] | 
0.01 mg  | 
| Standard repeatability [5% Max] | 
15 µg  | 
| Standard repeatability [Max] | 
20 µg  | 
| Permissible repeatability | 
35 µg  | 
| Linearity | 
±0.12 mg  | 
| Eccentricity (tested load) | 
0 mg  | 
| Electric compensation range | 
-10 g – +50 g  | 
| Stabilization time | 
30 s  | 
| Adjustment | 
internal (automatic)  | 
| Display | 
10″ graphic colour touchscreen  | 
| Weighing pan dimensions | 
ø75 mm  | 
| Weighing device dimensions | 
350×405×650  mm  | 
| Controling device dimensions | 
249×170×72 mm  | 
| Anti-draft chamber dimensions | 
660×470×700 mm  | 
| Packaging dimensions W x D x H | 
1000×900×685 mm  | 
| Chamber packaging size | 
930×750×1000  mm  | 
| Net weight | 
71 kg  | 
| Gross weight | 
98 kg  | 
| Supplementary weights | 
internal (automatic)  | 
| Supplementary weights internal | 
half automatic  | 
| Supplementary weights external | 
-  | 
| Communication interface | 
2×USB-A, USB-C, RS 232 (COM3), HDMI, Ethernet, Wi-Fi®, Hotspot  | 
| Power supply | 
100 – 240 V AC 50/60 Hz  | 
| Operating temperature | 
+15 – +30 °C  | 
| Operating  temperature change rate | 
±0.5 °C / 12 h (±0.3 °C / 4 h)  | 
| Relative humidity | 
40% – 60%  | 
| Relative humidity change rate | 
±5% / 12 h (3% / 4 h)  |